首页 | 本学科首页   官方微博 | 高级检索  
     


Simultaneous determination of trace aluminum (III), copper (II) and cadmium (II) in water samples by square-wave adsorptive cathodic stripping voltammetry in the presence of oxine
Authors:A. M. Beltagi  M. M. Ghoneim
Affiliation:(1) Chemistry & Physics Department, Faculty of Education, Kafr El-Sheikh University, 33616 Kafr El-Sheikh, Egypt;(2) Analytical & Electrochemistry Research Unit, Chemistry Department, Faculty of Science, Tanta University, 31527 Tanta, Egypt
Abstract:A validated adsorptive cathodic stripping voltammetry method is described for simultaneous determination of Al(III), Cu(II) and Cd(II) in water samples. In acetate buffer (pH 5) containing 10 μM oxine, these metal ions were determined as oxine complexes following adsorptive accumulation onto the HMDE at −0.05 V versus Ag/AgCl/KCls. The best signal to noise ratio was obtained using a square wave of scan increment 10 mV, frequency 120 Hz, and pulse-amplitude 25 mV. Limits of detection as low as 0.020 μg L−1 Al(III), 0.012 μg L−1 Cu(II) and 0.028 μg L−1 Cd(II) were achieved. Interference due to various cations (K(I), Na(I), Mg(II), Ca(II), Mn(II), Fe(III), Bi(III), Sb(III), Se(IV), Pb(II), Zn(II), Ni(II), Co(II)), anions (Cl, NO3−, SO4 2−, PO4 3−) and ascorbic acid was minimal as the measured signals change by 4% at the maximum. The stripping voltammetry method was successfully applied for simultaneous determination of Al(III), Cu(II) and Cd(II) in tap and natural bottled water samples.
Keywords:Aluminum  Copper  Cadmium  Oxine  Determination  Square-wave adsorptive stripping voltammetry
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号