首页 | 本学科首页   官方微博 | 高级检索  
     


Signal-to-noise measurements utilizing a novel dual-energymultimedia detector
Authors:Giakos   G.C. Chowdhury   S. Shah   N. Vedantham   S. Passerini   A.G. Suryanarayanan   S. Nemer   R. Mehta   K. Patnekar   N. Sumrain   S. Nataraj   K. Scheiber   C.
Affiliation:Dept. of Biomed. Eng., Akron Univ., OH;
Abstract:Dual-energy measurements are presented utilizing a novel slot-scan digital radiographic imaging detector, operating on gaseous solid state ionization principles. The novel multimedia detector has two basic functional components: a noble gas-filled detector volume operating on gas microstrip principles, and a solid state detector volume. The purpose of this study is to investigate the potential use of this multimedia detector for enhanced dual-energy imaging. The experimental results indicate that the multimedia detector exhibits a large subtracted signal-to-noise ratio. Although the intrinsic merit of this device is being explored for medical imaging, potential applications of the multimedia detector technology In other industrial areas, such as aerospace imaging, aviation security, and surveillance, are also very promising