首页 | 本学科首页   官方微博 | 高级检索  
     


Monochromator-induced glitches in EXAFS data I. Test of the model for linearly tapered samples
Authors:F. Bridges   G. G. Li  Xun Wang
Affiliation:

University of California, Santa Cruz, CA 95064, USA

Abstract:When EXAFS samples are slightly nonuniform in thickness, spikes, referred to as “glitches”, are observed in the data. Such features are intrinsic to data collected using crystal monochromators. We have recently developed a model to show why such features, persist in ratioed data. Here we test the predictions for linearly tapered samples. Our results suggest that the glitch mplitude can be significantly reduced by taking a second spectrum with the sample inverted, and then averaging the two spectra.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号