Design and construct of a new detector for gas chromatography based on continuous negative corona discharge
Authors:Ghahfarokhi M Sharifian  Khayamian T
Affiliation:Department of Chemistry, Isfahan University of Technology, Isfahan 84156-83111, Iran.
Abstract:In this work, a new detector was designed and constructed based on negative corona discharge. This detector can be used separately or as a detector in gas chromatography. The detector and chromatographic variables including cell temperature, gas flow rates, voltage between the two electrodes, and column temperature were optimized. Chloroform was used as a test compound to evaluate the performance of the detector. The detection limit of chloroform was obtained 0.78 ng∕ml and its dynamic range was over the range of 2-840 ng∕ml. The relative standard detection was about 6% for the limit of quantification. This detector is able to be used as an alternative for analysis of compounds containing electronegative elements.
Keywords:
本文献已被 PubMed 等数据库收录!