The mouse chromosome 7 distal imprinting domain maps to G-bands F4/F5 |
| |
Authors: | CV Beechey ST Ball KM Townsend J Jones |
| |
Affiliation: | Department of Radiology, University of Pittsburgh, PA 15261-0001, USA. maitz@rad.arad.upmc.edu |
| |
Abstract: | RATIONALE AND OBJECTIVES: The authors designed, assembled, tested, and clinically evaluated a high-quality, fast, and relatively inexpensive telemammography system. METHODS: The authors designed a telemammography system that uses a high-resolution film digitizer and high data compression (> or = 40:1) to send images over regular telephone lines to a high-resolution laser printer that produces images with the look and feel of the original image and can operate in a hub and spokes mode. The authors then evaluated the system's performance. In a preliminary clinical study, interpretations of the laser-printed system's output of 119 cases were compared with the original interpretations, followed by a review of any clinically significant differences. RESULTS: With the exception of the laser printer, which is a modified off-the-shelf product, all hardware components of the system are commercially available products. The system digitizes (50 microns pixel size), compresses, transmits, receives, decompresses, and prints a 30 MB mammography file in less than 4 minutes. In the clinical study, there were 13 differences (in 13 cases) in the level of concern or recommendations. Seven were found to be clinically insignificant by a third-party review. The remaining six were reviewed by the original interpreter, and three were determined to be significant enough for further action. All were found to result from intra-reader variability rather than differences in visualization of possible abnormalities. CONCLUSIONS: Almost real-time, high-quality telemammography without geographic boundaries is possible with the use of high-level data compression. Telemammography with laser-printed film as the display may make it possible to offer mammographic services in remote locations while using commercially available technology. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|