首页 | 本学科首页   官方微博 | 高级检索  
     


Status and research programs of the multinuclide accelerator mass spectrometry system at the University of Tsukuba
Authors:Kimikazu Sasa  Tsutomu Takahashi  Yuki Tosaki  Yuki Matsushi  Keisuke Sueki  Michiko Tamari  Takahiro Amano  Toshiyuki Oki  Shozo Mihara  Yoshihiro Yamato  Yasuo Nagashima  Kotaro Bessho  Norikazu Kinoshita  Hiroshi Matsumura
Affiliation:1. AMS Group, Tandem Accelerator Complex, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki 305-8577, Japan;2. Micro Analysis Laboratory, Tandem Accelerator, The University of Tokyo, 2-11-16 Yayoi, Bunkyo-ku, Tokyo 113-0032, Japan;3. Radiation Science Center, High Energy Accelerator Research Organization, 1-1 Oho, Tsukuba, Ibaraki 305-0801, Japan
Abstract:We present the current status and research programs of a multinuclide accelerator mass spectrometry (AMS) system on the 12UD Pelletron tandem accelerator at the University of Tsukuba (Tsukuba AMS system), Japan. A maximum terminal voltage of 12 MV is available for the AMS system. The Tsukuba AMS system can measure environmental levels of long-lived radioisotopes of 14C, 26Al, 36Cl and 129I by employing a molecular pilot beam. Recently, enhancements in AMS techniques and equipment, including sample preparation, the ion source and the data acquisition system, have improved the performance of 36Cl-AMS. The standard deviation of fluctuations is typically ±2%, and the machine background level for the 36Cl/Cl ratio is lower than 1 × 10?15 with a halite sample. We have measured over 500 samples in 1 year, including samples for earth and environmental sciences and nuclear safety research.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号