首页 | 本学科首页   官方微博 | 高级检索  
     


Boron suppression with a gas ionization chamber at very low energies (E < 1 MeV)
Authors:AM Müller  M Christl  M Döbeli  PW Kubik  M Suter  H-A Synal
Affiliation:Ion Beam Physics, Paul Scherrer Institute and ETH-Zurich, 8093 Zurich, Switzerland
Abstract:Efficient boron suppression for precise 10Be measurements with AMS is crucial. The performance of ΔE ? Eres gas ionization chambers is also very important for isobar suppression at low beam energies (<1 MeV). A boron suppression of 6–7 orders of magnitude is achievable with the ETH ΔE ? Eres gas ionization in the standard operational mode (readout of ΔE and Eres electrodes). Some physical effects such as pulse height defect or the energy focusing effect within a ΔE section will be discussed, emphasizing 10Be measurements below 1 MeV. Additionally the potential of silicon nitride membranes as passive energy degraders in front of the detector is demonstrated.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号