首页 | 本学科首页   官方微博 | 高级检索  
     

主动式双波长红外激光测温标定实验研究
引用本文:曲岩,宦克为,安保林,董伟,赵云龙,宋旭尧,原遵东.主动式双波长红外激光测温标定实验研究[J].计量学报,2021,42(2):137-143.
作者姓名:曲岩  宦克为  安保林  董伟  赵云龙  宋旭尧  原遵东
作者单位:长春理工大学,吉林长春130022;中国计量科学研究院,北京100029
基金项目:国家自然科学基金;吉林省科技发展计划项目
摘    要:基于主动式双波长红外激光的测温方法,可实现对未知发射率表面真实温度的测量.高精度的标定源是保证双波长红外测温系统测温精确性的重要基础,但目前双波长测温领域的研究工作缺乏关于标定实验结果的公开报道.因此,设计搭建了主动式红外激光测温标定源,研究该标定源的稳定性和均匀性,并对双波长激光测温系统进行标定.结果表明:所搭建主动...

关 键 词:计量学  双波长红外激光测温法  辐射测温  发射率  标定
收稿时间:2020-10-30

Study on Active Dual-wavelength Unfrared Laser Thermometry Calibration Experiments
QU Yan,HUAN Ke-wei,AN Bao-lin,DONG Wei,ZHAO Yun-long,SONG Xu-yao,YUAN Zun-dong.Study on Active Dual-wavelength Unfrared Laser Thermometry Calibration Experiments[J].Acta Metrologica Sinica,2021,42(2):137-143.
Authors:QU Yan  HUAN Ke-wei  AN Bao-lin  DONG Wei  ZHAO Yun-long  SONG Xu-yao  YUAN Zun-dong
Affiliation:1. Changchun University of Science and Technology, Changchun, Jilin 130022, China
2. National Institute of Metrology, Beijing 100029, China
Abstract:Based on active dual-wavelength infrared laser thermometry method, it is possible to measure the true surface temperature when the emissivity is unknown. A high-precision calibration source is an important basis for the accuracy of dual-wavelength infrared thermometry system. However, there is a lack of public reports on such calibration sources in the current research work in the field of dual-wavelength temperature measurement. Therefore, an active infrared laser temperature measuring source is designed and built to study the stability and uniformity of the calibration source, and the dual-wavelength laser thermometry system is calibrated. The results show that the active dual-wavelength infrared laser temperature calibration source has good stability, and the maximum temperature deviation within 20 minutes at 1173K is 0.22K; the surface temperature uniformity is well, and the standard deviation of the surface temperature at 1173K is 0.34K. When the true temperature of the source surface is above 923K, the relative standard deviation of the collected signal is less than 0.7%. The calibration experiment results show that the calibration source has good reliability for the accurate calibration of the active dual-wavelength infrared laser temperature measurement system.
Keywords:metrology  dual-wavelength infrared laser thermometry method  radiation thermometry  emissivity  calibration  
本文献已被 CNKI 万方数据 等数据库收录!
点击此处可从《计量学报》浏览原始摘要信息
点击此处可从《计量学报》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号