首页 | 本学科首页   官方微博 | 高级检索  
     

三段控温固定点炉复现铝凝固点研究
引用本文:何沛,林林,闫小克,张明宇,王宁. 三段控温固定点炉复现铝凝固点研究[J]. 计量学报, 2021, 42(3): 321-326. DOI: 10.3969/j.issn.1000-1158.2021.03.10
作者姓名:何沛  林林  闫小克  张明宇  王宁
作者单位:北京科技大学,北京100083;中国计量科学研究院,北京100029;北京科技大学,北京100083;中国计量科学研究院,北京100029
摘    要:固定点炉垂直温场均匀性是影响ITS-90国际温标固定点温坪质量的重要因素.为了提高铝凝固温坪的复现水平,设计了三段控温固定点炉,利用金属外壳铝固定点容器,研究了垂直温场均匀性及其影响因素;在此基础上,采用连续热流密度法高精度复现了铝凝固点.实验结果表明:通过调整上部、中部、下部炉温的设置,可改善固定点炉温场均匀性,铝凝...

关 键 词:计量学  铝凝固点  三段控温  固定点炉  垂直温场  凝固温坪  ITS-90
收稿时间:2020-10-09

Study on the Realization of Aluminum Freezing Point in the Three-zone Temperature Controlled Fixed Point Furnace
HE Pei,LIN Lin,YAN Xiao-ke,ZHANG Ming-yu,WANG Ning. Study on the Realization of Aluminum Freezing Point in the Three-zone Temperature Controlled Fixed Point Furnace[J]. Acta Metrologica Sinica, 2021, 42(3): 321-326. DOI: 10.3969/j.issn.1000-1158.2021.03.10
Authors:HE Pei  LIN Lin  YAN Xiao-ke  ZHANG Ming-yu  WANG Ning
Affiliation:1. University of Science and Technology Beijing, Beijing 100083, China
2. National Institute of Metrology, Beijing 100029, China
Abstract:The vertical temperature uniformity of the fixed-point furnace is an important factor affecting the plateau quality of the fixed-points of the International Temperature Scale of 1990 (ITS-90). In order to improve the reproducibility of the aluminum freezing point, a three-zone temperature-controlled fixed-point furnace was designed, and the aluminum fixed point cell with a metal shell was used to study the vertical temperature uniformity and its influencing factors. On this basis, the continuous heat flux method was adopted to precisely realize the aluminum freezing point. The experimental results show that the temperature uniformity can be improved by adjusting the settings of the top, middle, and bottom furnace temperatures, and the temperature stability of the freezing plateau within 4 hours is less than 1mK. The phase transition supercooling of the high purity aluminum is approximately 0.14℃ under the experimental conditions.
Keywords:metrology  aluminum freezing point  three-zone temperature controlled  fixed point furnace  vertical temperature uniformity  freezing plateau  ITS-90  
本文献已被 CNKI 万方数据 等数据库收录!
点击此处可从《计量学报》浏览原始摘要信息
点击此处可从《计量学报》下载全文