首页 | 本学科首页   官方微博 | 高级检索  
     

微型双温度固定点容器研制
引用本文:曾佳旭,潘江,孙建平,杨月,朱天梦,王成科.微型双温度固定点容器研制[J].计量学报,2021,42(4):458-462.
作者姓名:曾佳旭  潘江  孙建平  杨月  朱天梦  王成科
作者单位:中国计量大学,浙江杭州310018;中国计量科学研究院,北京100029
基金项目:国家重点研发计划重点专项
摘    要:基准固定点传递技术应用于现场温度校准已成为提高工业温度测量水平的一种重要途径。采用多孔石墨坩埚半包围结构,内部对称填充了高纯铟(In)和锡(Sn)2种金属,研制了一种应用于现场校准的微型双温度固定点容器。实验结果表明,In点熔化温坪持续时间约为2 h,Sn点熔化温坪持续时间约为3 h,In和Sn的温坪复现的扩展不确定度分别为4.0 mK 和4.4 mK(k=2),可满足工业现场对精密铂电阻温度计的校准需求。

关 键 词:计量学  微型固定点容器  铟点  锡点  现场校准  铂电阻温度计
收稿时间:2019-09-20

Development of Miniature Double Temperature Fixed Point Cell
ZENG Jia-xu,PAN Jiang,SUN Jian-ping,YANG Yue,ZHU Tian-meng,WANG Cheng-ke.Development of Miniature Double Temperature Fixed Point Cell[J].Acta Metrologica Sinica,2021,42(4):458-462.
Authors:ZENG Jia-xu  PAN Jiang  SUN Jian-ping  YANG Yue  ZHU Tian-meng  WANG Cheng-ke
Affiliation:1. China Jiliang University, Hangzhou, Zhejiang 310018, China
2. National Institute of Metrology, Beijing 100029, China
Abstract:Applying the fixed point transfer technology to on-site temperature calibration has become an important way to improve the industrial temperature measurement level. A miniature dual temperature fixed point for field calibration was developed. The fixed point adopted a porous graphite crucible and a semi-enclosed structure filled with two metals, indium and tin. Experimental results showed, at indium point, the temperature plateau curve lasted about 2 hours, and at tin point, the temperature plateau curve lasted about 3 hours. The extended uncertainty reproduced in temperature plateau curve of In was 4.0mK (k=2), and Sn was 4.4mK (k=2). Therefore, the miniature double temperature fixed point device could meet the calibration requirements of precision platinum resistance thermometers in industrial field.
Keywords:metrology  mini fixed point cell  indium point  tin point  field calibration  platinum resistance thermometer  
本文献已被 CNKI 万方数据 等数据库收录!
点击此处可从《计量学报》浏览原始摘要信息
点击此处可从《计量学报》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号