首页 | 本学科首页   官方微博 | 高级检索  
     


Understanding TES microcalorimeter noise and energy resolution
Authors:Wouter Bergmann Tiest  Marcel Bruijn  Henk Hoevers  Piet de Korte  Jan van der Kuur  Wim Mels
Affiliation:

SRON National Institute for Space Research, Sorbonnelaan 2, Utrecht 3584 CA, The Netherlands

Abstract:We have performed numerical calculations of the noise in voltage-biased superconducting transition edge-based X-ray microcalorimeters, using a finite-element model. Details of the model are discussed, as well as results for different absorber geometries. The results are in agreement with an analytical model and show that the amount of internal thermal fluctuation noise can be reduced by using a segmented absorber. The simulation also agrees well with experimental data, which, for our detectors, contain no major unidentified noise sources. Furthermore, the discrepancy between the (small-signal) theoretical and the measured energy resolutions for 5.9 keV X-rays, for our sensors typically a factor of not, vert, similar2, is explained by a more accurate modelling of the sensor responsivity, taking into account the large signal behaviour.
Keywords:
本文献已被 ScienceDirect 等数据库收录!