首页 | 本学科首页   官方微博 | 高级检索  
     


The content of 250Cf and 248Cm in 252Cf neutron sources and the effect on the neutron emission rate
Authors:Roberts N J  Jones L N
Affiliation:Neutron Metrology Group, DQL, National Physical Laboratory, Hampton Road, Teddington, Middlesex TW11 0LW, UK.
Abstract:One of the most common radionuclide neutron sources used for the calibration of detectors is (252)Cf. However, these sources also contain (250)Cf, which is present in the material from which the sources are made, and (248)Cm, which is formed as the daughter of (252)Cf via alpha-decay. Both decay by spontaneous fission with longer half-lives than (252)Cf. Consequently, as the source becomes older, the emission rate does not follow the decay curve of (252)Cf. Fits have been made to emission rate measurements of (252)Cf sources at NPL spanning over 30 y to deduce their (250)Cf and (248)Cm content. The emission rate of a source can be significantly underestimated if the presence of (250)Cf and (248)Cm is not taken into account, and this has been investigated for a typical (252)Cf source. The importance of this problem to other calibration laboratories and users of (252)Cf sources is emphasised.
Keywords:
本文献已被 PubMed 等数据库收录!