首页 | 本学科首页   官方微博 | 高级检索  
     


Slurry analysis after lead collection on a sorbent and its determination by electrothermal atomic absorption spectrometry
Authors:Baysal Asli  Tokman Nilgun  Akman Suleyman  Ozeroglu Cemal
Affiliation:

aIstanbul Technical University, Faculty of Science and Letters, Department of Chemistry, 34469 Maslak-Istanbul, Turkey

bIstanbul University, Department of Chemistry, Faculty of Engineering, 34320 Avcilar-Istanbul, Turkey

Abstract:In this study, in order to eliminate the drawbacks of elution step and to reach higher enrichment factors, a novel preconcentration/separation technique for the slurry analysis of sorbent loaded with lead prior to its determination by electrothermal atomic absorption spectrometry was described. For this purpose, at first, lead was collected on ethylene glycol dimethacrylate methacrylic acid copolymer (EGDMA-MA) treated with ammonium pyrolidine dithiocarbamate (APDC) by conventional batch technique. After separation of liquid phase, slurry of the sorbent was prepared and directly pipetted into graphite furnace of atomic absorption spectrophotometer. Optimum conditions for quantitative sorption and preparation of the slurry were investigated. A 100-fold enrichment factor could be easily reached.

The analyte element in certified sea-water and Bovine-liver samples was determined in the range of 95% confidence level. The proposed technique was fast and simple and the risks of contamination and analyte loss were low. Detection limit (3δ) for Pb was 1.67 μg l−1.

Keywords:Separation  Preconcentration  Lead  Electrothermal atomic absorption spectrometry  Slurry sampling
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号