首页 | 本学科首页   官方微博 | 高级检索  
     


A tester for artificial respirators
Authors:G Belforte  G Eula  T Raparelli
Affiliation:

Department of Mechanics, Politecnico di Torino, Technological University, C.so Duca degli Abruzzi, 24, 10129 Torino, Italy

Abstract:All equipment used for resuscitation must be validated before use. This is possible only if means are available for simulating the human airways and lungs, reproducing the conditions in which the respirators will be employed. It is also important that such simulators be capable of determining all variables (flow rate, respiratory rate, pressure, etc.) that can be regulated with an artificial ventilator in full agreement with international standards. This paper illustrates the development of a lightweight, portable prototype lung model with a compliant bellows whereby various lung conditions can be simulated. The prototype was validated using a fixed-capacity reference model which ensures good measurement accuracy, though its weight and bulk are by no means inconsiderable. The investigation was carried out by sizing each component in accordance with international and American standards.
Keywords:Pulmonary apparatus model  Human lung prototype
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号