首页 | 本学科首页   官方微博 | 高级检索  
     


Development and validation of a high-performance liquid chromatography-mass spectrometry assay for determination of amphetamine, methamphetamine, and methylenedioxy derivatives in meconium
Authors:Pichini Simona  Pacifici Roberta  Pellegrini Manuela  Marchei Emilia  Lozano Jaime  Murillo Janeth  Vall Oriol  García-Algar Oscar
Affiliation:Drug Research and Control Department, Istituto Superiore di Sanità, Viale Regina Elena 299, 00161 Rome, Italy. pichini@iss.it
Abstract:A procedure based on liquid chromatography-mass spectrometry (LC-MS) is described for determination of amphetamine, methamphetamine, and methylendioxy derivatives in meconium, using 3,4-methylendioxypropylamphetamine as internal standard. The analytes were initially extracted from the matrix by 17 mM methanolic HCl. Subsequently, a solid-phase extraction with Bondelut Certify columns was applied. Chromatography was performed on a C(18) reversed-phase column using a linear gradient of 10 mM ammonium bicarbonate, pH 9.0-methanol as a mobile phase. Analytes were determined in LC-MS single ion monitoring mode with an atmospheric pressure ionization-electrospray interface. The method was validated in the range 0.005-1.00 microg/g using 1 g of meconium per assay. Mean recoveries ranged between 61.1 and 87.2% for different analytes. The quantification limits were 0.005 microg/g meconium for amphetamine, methamphetamine, and 4-hydroxy-3-methoxymethamphetamine and 0.004 microg/g meconium for 3,4-methylenedioxyamphetamine, 3,4-methylenedioxymethamphetamine, 3,4-methylenedioxyethylamphetamine, and N-methyl-1-(3,4-methylenedioxyphenyl)-2-butanamine. The method was applied to analysis of meconium in newborns to assess eventual fetal exposure to amphetamine derivatives.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号