Common-path optical coherence tomography guided fibre probe for spatially precise optical nerve stimulation |
| |
Authors: | Zhang K Katz E Kim DH Kang JU Ilev IK |
| |
Affiliation: | Department of Electrical and Computer Engineering, The Johns Hopkins University, 3400 North Charles Street, Baltimore, MD, USA; |
| |
Abstract: | A simple and effective common-path optical coherence tomography guided fibre probe for optical nerve stimulation has been developed and tested. The probe is capable of real-time monitoring of the fibre-to-nerve distance up to more than 1 mm at an axial resolution of approximately 3 μm, thus improving the precision and safety of stimulation laser power delivery. |
| |
Keywords: | |
|
|