首页 | 本学科首页   官方微博 | 高级检索  
     

220Rn测量仪器的刻度问题
引用本文:丘寿康. 220Rn测量仪器的刻度问题[J]. 核技术, 2003, 26(7): 561-563
作者姓名:丘寿康
作者单位:南华大学核科学技术学院,衡阳,421001
摘    要:调研了各类钍射气(^220Rn)测量仪的刻度方法和存在的问题。针对各类^220Rn标准源的析出百分率常小于100%而有待准确测定这一关键问题,建立了流气闪烁室绝对测量方法,并用活性炭-γ谱法作了验证,解决了准确测定^220Rn标准源析出率的问题,为提高^220Rn测量仪刻度的准确性创造了条件。

关 键 词:220Rn测量  刻度  流气法
修稿时间:2001-08-13

Calibration of thoron monitors
QIU Shoukang. Calibration of thoron monitors[J]. Nuclear Techniques, 2003, 26(7): 561-563
Authors:QIU Shoukang
Abstract:The calibration of various kinds of thoron monitor has been investigated. A main problem lies in that the emanation percentage of various kinds of thoron standard source is always less than 100%. In order to determine the emanation percentage accurately, an absolute measurement method with airflow-through scintillation cell was established, and its feasibility was verified by a comparative measurement with the activated charcoal-g spectrometer method. This will make the calibration accuracy of thoron monitors rising.
Keywords:Thoron   Measurement   Calibration   Airflow method CLC O613.16  
本文献已被 CNKI 维普 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号