首页 | 本学科首页   官方微博 | 高级检索  
     


Semi-automatic measurement of rigid gas-permeable contact lens movement in keratoconus patients using blinking images
Authors:Hassan Rezazadeh  Hossein Arabalibeik  Fateme Alipour  Pezhman Pasyar  Vahid Sadeghi
Affiliation:1. Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran;2. Research Center of Biomedical Technology and Robotics (RCBTR), Tehran University of Medical Sciences, Tehran, Iran;3. Eye Research Center, Farabi Eye Hospital, Tehran University of Medical Sciences, Tehran, Iran
Abstract:

Purpose

To introduce a method for estimation of the rigid gas-permeable contact lens (RGP) movement.

Materials and methods

Videos captured from normal blinking of keratoconus patients while wearing RGP lenses were used for this study. The videos are recorded using the CCD camera of a smart phone attached to the eyepiece of the slit lamp. The algorithm starts with extracting two frames of the video related to the highest and lowest positions of the lens during blinking, followed by an appropriate edge detection method. In the next step circular Hough transform is used to find the center of lens and to segment it in each image. Finally the lens movement is estimated by measuring vertical displacement of the lens center between these two frames.

Results

Mean and standard deviation of the difference between real movement and results of the algorithm for 20 cases are ?8.66% and 10.71% respectively. The results are highly correlated with Pearson coefficient 0.986?P?<?0.001. Bland-Altman plot with 95% levels of agreement (LoA) shows an agreement between exact manual measurement method and the proposed algorithm.

Conclusion

The proposed algorithm shows a relatively high accuracy as the first attempt and compared to the routine qualitative visual estimation. Considering the importance of the lens movement, although this system was not tested on a series of RGP fitting patients yet, semi-automatic measurement may potentially help practitioners decide the appropriate RGP lens fit and reduce the fitting time.
Keywords:Circular Hough transform  Fluorescein pattern  Keratoconus  Lens movement  RGP contact lens
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号