首页 | 本学科首页   官方微博 | 高级检索  
     


Field instrument for simultaneous large dynamic range measurement of atmospheric hydrogen sulfide, methanethiol, and sulfur dioxide
Authors:Toda Kei  Ohira Shin-Ichi  Tanaka Takayoshi  Nishimura Tomohiko  Dasgupta Purnendu K
Affiliation:Department of Environmental Science, Faculty of Science, Kumamoto University, 2-39-1 Kurokami, Kumamoto 860-8555, Japan. todakei@sci.kumamoto-u.ac.jp
Abstract:We describe a membrane-based collection/analysis system that differentially monitors H2S and CH3SH, and to which a conductometric SO2 analyzer using the same collector was coupled. A diffusion scrubber (DS) comprised of a Nafion tube collects H2S selectively while a porous polytetrafluoroethylene (pPTFE) DS collects both H2S and CH3SH. Both gases are measured via their ability to react with fluorescein mercuric acetate (FMA) which results in decreased fluorescence. The limited dynamic range of a negative signal procedure was overcome by using dual DS units comprised of short and long scrubbers, placed serially in the liquid flow line. Different DS designs and membrane materials were investigated. H2S, CH3SH, and SO2 from a biogenic point source were continuously measured, and the H2S/CH3SH data compared well with a standard procedure involving Tedlar bag collection, preconcentration and thermal desorption from a Tenax trap, and measurement by gas chromatography/flame photometric detection. Walkaround portability of the instrument and very large dynamic range measurement of H2S and SO2 were demonstrated around the Mt. Aso volcano.
Keywords:
本文献已被 PubMed 等数据库收录!