首页 | 本学科首页   官方微博 | 高级检索  
     


Blood triggered corrosion of magnesium alloys
Authors:J Geis-Gerstorfer  Ch SchilleE Schweizer  F RuppL Scheideler  H-P ReichelN Hort  A NolteH-P Wendel
Affiliation:a Section Medical Materials & Technology, University Hospital Tuebingen, Tuebingen, Germany
b Weisensee Company, Eichenzell, Germany
c GKSS Research Center, Institute of Materials Research, Geesthacht, Germany
d Div. Congenital & Paediatric Cardiac Surgery, University Children''s Hospital, Tuebingen, Germany
Abstract:Intravascular stents manufactured out of bioabsorbable magnesium (Mg) or Mg-alloys are considered as auspicious candidates for the next stent generation. However, before clinical application numerous physical and biological tests, especially to predict the clinically highly important degradation kinetics in vivo, have to be performed. In a Chandler-Loop model, the initial degradation of eight different magnesium alloys during 6 h in contact with human whole blood was investigated. The magnesium release varied between 0.91 ± 0.33 mg/cm2 (MgAl9Zn1) and 2.57 ± 0.38 mg/cm2 (MgZn1). No correlation could be found with Mg release data obtained after immersion in simulated body fluid (SBF). This pilot study showed that Mg corrosion is highly influenced by the biological test environment (SBF or blood, etc.) and that a modified Chandler-Loop model with human whole blood may be superior to predict corrosion of Mg alloys under clinical conditions than the SBF models presently used.
Keywords:Stent  Mg release  Chandler-Loop  Static immersion test  Blood corrosion  Simulated body fluid (SBF)
本文献已被 ScienceDirect 等数据库收录!