首页 | 本学科首页   官方微博 | 高级检索  
     


An Electrodynamic Pulsed Valve for Molecular Beams
Authors:M. I. Mazurenko  V. A. Petukhov  V. A. Povedailo  M. A. Semenov  I. V. Smagin
Affiliation:(1) Lebedev Physical Institute, Russian Academy of Sciences, Leninskii pr. 53, Moscow, 117924, Russia;(2) Institute of Molecular and Atomic Physics, Belorussian Academy of Sciences, pr. F. Skoriny 70, Minsk, 220072, Belarus
Abstract:The pulsed valve with an electromagnetic drive, based on a permanent magnet and a movable coil, is intended to obtain pulsed molecular beams. It is controlled by small currents of a few amperes and, thus, does not produce considerable electrodynamic disturbances. An electric control circuit of the valve provides for reduced mechanical stress on the movable coil. The frequency of the valve actuation is controlled in a range of 5–50 Hz; the gas pulse duration is regulated within the limits of 0.2–4 ms (FWHM). This device is suggested for application in the spectroscopy of complex molecules in molecular beams.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号