Radiation transport calculations for 50 MV photon therapy beam using the Monte Carlo code GEANT4
Authors:Larsson Susanne  Svensson Roger  Gudowska Irena  Ivanchenko Vladimir  Brahme Anders
Affiliation:Department of Medical Radiation Physics, Karolinska Institutet and Stockholm University, P.O. Box 260, S-171 76 Stockholm, Sweden. Susanne.Larsson@radfys.ki.se
Abstract:A new thin transmission target technique for fast dose delivery using narrow scanned photon beams has been developed. High-energy, 50-100 MeV, electron beams of low emittance incident on thin low-Z targets produce narrow and intense high-energy bremsstrahlung beams. However, electrons transmitted through the target are bent from the therapeutic beam by a purging magnet and have to be effectively absorbed in a dedicated electron collector. The electron-photon transport through a treatment head has been studied using the Monte Carlo simulation toolkit Geant4. The Geant4 electromagnetic physics processes have been compared with experimental data of radial dose profiles. The differences between calculated and measured radial dose distributions are approximately 2-10%. Preliminary investigations of the collector design have been carried out in order to minimise secondary electron and photon contamination of the therapeutic beam. The toolkit presented here is promising for further development of narrow photon beam therapy.
Keywords:
本文献已被 PubMed 等数据库收录!