首页 | 本学科首页   官方微博 | 高级检索  
     


Pulsed laser damage thresholds in vitro for intraocular lenses and membranes
Authors:Davi  S Gaasterland  D Cummins  C  III Liesegang  G
Affiliation:National Institutes of Health, Bethesda, MD, USA;
Abstract:Energy and power density damage thresholds were determined in air, for plastic IOL's and membranes at the focal point of several solid-state laser systems: 1) 694 nm,Q-switched single pulse (30 ns), multimode, 2) 1064 nm,Q-switched single pulse (20 ns), TEM00, 3) 1060 nm, mode-locked single pulse, 15 ps, TEM11, 4) 530 nm, mode-locked single pulse, 15 ps, TEM11, and 5) 1064 nm, mode-locked pulse train (9-11 pulses, 30 ps), TEM00. Pulse energies bracketing damage thresholds as well as focal diameter and pulse duration for each system were determined. Energy density thresholds are lower, and power density thresholds higher, for shorter duration pulses-e.g., 23 J/cm2(1.15 GW/cm2) versus 6 J/cm2(400 GW/cm2) at the same wavelength as in systems 2) and 3) (p = 0.005). Damage thresholds for glass IOL's are 37 J/cm2(1.9 GW/cm2) and 37 J/cm2(1235 GW/cm2) as in systems 2) and 5). Damage threshold values for plastic membranes (Saran Wrap®) exposed to nanosecond and picosecond pulse trains of Nd:YAG at 1064 nm are about half that of plastic IOL's. When laser pulses with a cone angle of 14° from systems 2) and 5) are focused on plastic membrane next to the IOL, damage thresholds are 30 J/cm2(1.5 GW/cm2) for 20 nsQ-switched pulses and 20 J/cm2(670 GW/cm2) for trains of 30 ps mode-locked pulses. Damage thresholds of IOL's immersed in 0.9 percent saline are approximately the same as those obtained in air.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏