Measurement of thermal conductivity of fluid using single and dual wire transient techniques |
| |
Authors: | Siddharth Komini Babu KS Praveen B Raja P Damodharan |
| |
Affiliation: | Indian Institute of Information Technology, Design and Manufacturing (IIITD&M) Kancheepuram, Chennai 600 127, India |
| |
Abstract: | A modified measurement device to measure thermal conductivity of fluids using transient hot-wire technique has been designed, developed, tested and presented in this paper. The equipment is designed such that the thermal conductivity could be measured using both single wire sensor of different length and dual wire sensor. The sensor, which is also a heater, is a platinum micro-wire of 50 μm diameter. The influence of wire length on the measurement of thermal conductivity of fluids is tested using two single wires of length 50 mm and 100 mm. The thermal conductivity is also measured using a dual hot wire arrangement; which is achieved by placing the 100 mm and 50 mm wires in a Wheatstone bridge with the 100 mm wire as the sensor and 50 mm wire as a compensation wire. The apparatus requires a 100 ml of test fluid to perform the experiment. The testing temperature of the test fluid during the experimentation can be suitably varied by the choice of heat exchange fluid used in the apparatus. Water is chosen as testing fluids for primary standards. When compared to single wires, the thermal conductivity of the fluids measured is consistent with dual-wire method with an uncertainty of ±0.25%. |
| |
Keywords: | Transient hot wire Platinum Thermal conductivity Dual wire method |
本文献已被 ScienceDirect 等数据库收录! |