首页 | 本学科首页   官方微博 | 高级检索  
     


Development of the LED measurement system of chromaticity coordinates correction
Authors:Nyen-Ts Chen  Chih-Wei Chen
Affiliation:National Pingtung University of Science and Technology, Department of Mechanical Engineering, Taiwan, ROC
Abstract:In this paper, we propose a procedure to adjust the chromaticity coordinates by using a few samples of the color measurement coordinate to do the error correction and choose the correction method of minimum color error difference average to be the procedure of measuring instruments, and to construct an efficient and predictable calibration procedure. This research adopts some engineering statistics methods, such as median and average methods, and corrects the experiments to obtain the chromaticity coordinates. The four calibration methods that we designed lower the average value of the color errors as 0.034. Moreover, the minimum average value of errors in the X-axis could be revised to 6.3% while the minimum average value of errors in the Y-axis 8.8%. The accuracy of best calibrated chromaticity coordinates could achieve to the percentage 71.1%.
Keywords:Chromaticity coordinates  Engineering statistics  Median  Mean value
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏