Evaluation of the Parallel Conductor Theory for Measuring Human Limb Blood Flow by Electrical Admittance Plethysmography |
| |
Authors: | Shimazu Hideaki Yamakoshi Ken-Ichi Togawa Tatsuo Fukuoka Masakazu Ito Hiroshi |
| |
Affiliation: | Research Institute of Applied Electricity, Hokkaido University; |
| |
Abstract: | We devised a compensation technique that measures the admittance change of a limb submerged in an electrolyte solution in a cylinder. Using this technique, we evaluated the accuracy of admittance plethysmography and the validity of the parallel-conductor model on which the theory of blood flow measurement by electrical admittance (or impedance) plethysmography is based. From a theoretical point of view, if a limb is regarded as a parallel-conductor model, the admittance change due to blood pooling following venous occlusion should disappear when the resistivity of the solution is equal to that of the blood. |
| |