首页 | 本学科首页   官方微博 | 高级检索  
     


Design and fabrication of a 40-MHz annular array transducer
Authors:Ketterling Jeffrey A  Aristizábal Orlando  Turnbull Daniel H  Lizzi Frederic L
Affiliation:Riverside Research Institute, New York, NY, USA. ketterling@rrinyc.org
Abstract:This paper investigates the feasibility of fabricating a five-ring, focused annular array transducer operating at 40 MHz. The active piezoelectric material of the transducer was a 9-microm thick polyvinylidene fluoride (PVDF) film. One side of the PVDF was metallized with gold and forms the ground plane of the transducer. The array pattern of the transducer and electrical traces to each annulus were formed on a copper-clad polyimide film. The PVDF and polyimide were bonded with a thin layer of epoxy, pressed into a spherically curved shape, then back filled with epoxy. A five-ring transducer with equal area elements and 100-microm kerfs between annuli was fabricated and tested. The transducer had a total aperture of 6 mm and a geometric focus of 12 mm. The pulse/echo response from a quartz plate located at the geometric focus, two-way insertion loss (IL), complex impedance, electrical crosstalk, and lateral beamwidth all were measured for each annulus. The complex impedance data from each element were used to perform electrical matching, and the measurements were repeated. After impedance matching; fc approximately equal to 36 MHz and -6-dB bandwidths ranged from 31 to 39%. The ILs for the matched annuli ranged from -28 to -38 dB.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏