首页 | 本学科首页   官方微博 | 高级检索  
     


Direct plasma analysis of drug compounds using monolithic column liquid chromatography and tandem mass spectrometry
Authors:Hsieh Yunsheng  Wang Ganfeng  Wang Yuguang  Chackalamannil Samuel  Korfmacher Walter A
Affiliation:Drug Metabolism and Pharmacokinetics Department, Schering-Plough Research Institute, Kenilworth, New Jersey 07033, USA. yunsheng.hsieh@spcorp.com
Abstract:A monolithic silica column high-performance liquid chromatography-tandem mass spectrometry (HPLC-MS/MS) method has been developed for the high-speed direct simultaneous determination of a drug discovery compound and its major circulating metabolite (M-72) in rat plasma. This methodology makes use of flow programming and an alkyl-bonded silica rod column for fast macromolecule removal and chromatographic separation without the need for significant sample preparation. The matrix ionization suppression effect on the monolithic column HPLC-MS/MS system was investigated using the postcolumn infusion technique. After 200 plasma injections on a 50 x 4.6 mm monolithic silica column, consistent column efficiency of close to 39,000 theoretical plates/m and reproducible retention times for the analytes were observed. The apparent on-column recoveries of 12 test compounds in rat plasma samples were greater than 90%. The proposed fast direct plasma injection method was tested over a 3-day period with the interday coefficient of variation less than 15% for both analytes.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏