首页 | 本学科首页   官方微博 | 高级检索  
     


Wear of moderately cross-linked polyethylene in fixed-bearing total knee replacements
Authors:Brockett Claire L  Jennings Louise M  Hardaker Catherine  Fisher John
Affiliation:Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, UK. C.L.Brockett@leeds.ac.uk
Abstract:Cross-linked polyethylene has been introduced into total joint replacement to improve wear resistance. Although the performance of highly cross-linked polyethylene is well documented clinically and experimentally for total hip replacements, the reduction in mechanical properties with increasing irradiation is of concern for application to total knee replacement. The aim of this study was to investigate the wear performance of a moderately cross-linked polyethylene material in a fixed-bearing total knee replacement. The study was conducted using two femoral geometries, a conventional cruciate-retaining femoral and a high-flexion femoral geometry. The femoral geometry appeared to have no effect on the wear of the knee replacement under standard gait conditions. A significant reduction in wear volume was measured with the moderately cross-linked polyethylene compared with the conventional polyethylene over a six-million-cycle wear study. This study indicates the use of a moderately cross-linked polyethylene in a fixed-bearing total knee replacement may provide a low wearing option for total knee replacement.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号