首页 | 本学科首页   官方微博 | 高级检索  
     


Comparison of different methods for shielding design in computed tomography
Authors:Ciraj-Bjelac O  Arandjic D  Kosutic D
Affiliation:Radiation and Environmental Department Protection, Vinca Institute of Nuclear Sciences, MP Alasa 12-14, Vinca, PO Box 522, 11001Belgrade, Serbia. ociraj@vinca.rs
Abstract:The purpose of this work is to compare different methods for shielding calculation in computed tomography (CT). The BIR-IPEM (British Institute of Radiology and Institute of Physics in Engineering in Medicine) and NCRP (National Council on Radiation Protection) method were used for shielding thickness calculation. Scattered dose levels and calculated barrier thickness were also compared with those obtained by scatter dose measurements in the vicinity of a dedicated CT unit. Minimal requirement for protective barriers based on BIR-IPEM method ranged between 1.1 and 1.4 mm of lead demonstrating underestimation of up to 20 % and overestimation of up to 30 % when compared with thicknesses based on measured dose levels. For NCRP method, calculated thicknesses were 33 % higher (27-42 %). BIR-IPEM methodology-based results were comparable with values based on scattered dose measurements, while results obtained using NCRP methodology demonstrated an overestimation of the minimal required barrier thickness.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号