首页 | 本学科首页   官方微博 | 高级检索  
     


A Method and a Device Prototype for Noninvasive Measurements of Blood Perfusion in a Tissue
Authors:D G Lapitan  O A Raznitsyn
Affiliation:1.Vladimirskiy Moscow Region Research Clinical Institute,Moscow,Russia;2.ООО EOS-Medika Research and Development Center,Moscow,Russia
Abstract:Noncoherent fluctuation spectroscopy (NFS) is described as an optical method for measuring the tissue blood perfusion. This method is based on the registration of low-frequency fluctuations of a backscattered signal from a tissue in the frequency range of 0–20 Hz. It allows the assessment of the blood-flow intensity in small blood vessels of the skin by calculating an integral parameter, the perfusion index, which is similar to that in the laser Doppler flowmetry (LDF) method. A device prototype in which LEDs at wavelengths of 568 and 808 nm are used in a miniature optical head was developed. The device makes it possible to perform diagnostics during the subject’s motion via the use of LED light sources. It is shown that the NFS signal is comparable with the LDF signal. Thus, the NFS method can serve as an inexpensive alternative to the LDF method.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏