The non-equilibrium response of a high-resolution Ti/Au X-ray microcalorimeter
Authors:S. Terracol    T. Miyazaki    U. Morita    Y. Ishisaki    K. Tanaka    K. Mitsuda   S. Friedrich
Affiliation:

a Lawrence Livermore National Laboratory, 7000 East Av, P.O. Box 808, L-270, Livermore, CA 94550, USA

b Tokyo Metropolitan University, Minami-Osawa 1-1, Hachioji-shi, Tokyo 192-0397, Japan

c Seiko Instruments Inc, 563 Takatsuka-Shinden, Matsudo-shi, Chiba 270-2222, Japan

d Institute of Space and Astronautical Science, 3-1-1 Yoshinodai, Sagamihara, Kanagowa 229-8810, Japan

Abstract:We have studied the response of a high resolution Ti/Au transition edge sensor (TES) microcalorimeter with an energy resolution of 6.2±0.7 eV at 1.5 keV and 7.8±0.9 eV at 6.4 keV. We find that the sensitivity ≡R/T dR/dT of the TES varies significantly along the superconducting-to-normal transition that the bias point traverses during a pulse. Furthermore, is reduced significantly during a pulse compared to its equilibrium value calculated from the DC characteristics. This leads to an overestimate of the expected energy resolution when basing the prediction on the equilibrium sensitivity.
Keywords:Author Keywords: X-ray microcalorimeter   Transition edge sensor   Non-equilibrium sensitivity