Abstract:A novel gas chromatography detector is described that uses acoustic signals from a partly premixed hydrogen-air flame burning on top of a capillary. The device, referred to as the acoustic flame detector (AFD), is based on the measurement of the frequency of acoustic transients generated at the burner under a range of operating conditions. The presence of trace amounts of analyte in the flame was found to increase the frequency of these sonic bursts from the baseline level of ~100 Hz. The response of the AFD for n-dodecane, measured as the shift in frequency, was determined to be linear over ~3 orders of magnitude, with a minimum detectable level of about 1-5 ng C/s using the current system. The sensitivity correlates roughly with carbon content, except for certain organometallics (Sn, Mn), which gave substantially enhanced signals. Some tailing was observed but became serious only for particular types of organometallics. The noise of the system was predominantly of the 1/f type. The effects of flow conditions, burner geometry, and flame gas constituents were investigated. The oscillations could be followed by acoustic, visual, electrical, and optical means. The AFD mechanism is shown to involve oscillatory chemical kinetics, in which the flame front (the inner cone) temporarily enters a few millimeters into the capillary during each cycle, thereby creating the acoustic signal.