首页 | 本学科首页   官方微博 | 高级检索  
     


High-pressure stopped-flow spectrometer for kinetic studies of fast reactions by absorbance and fluorescence detection
Authors:Bugnon P  Laurenczy G  Ducommun Y  Sauvageat P Y  Merbach A E  Ith R  Tschanz R  Doludda M  Bergbauer R  Grell E
Affiliation:Institut de Chimie Minérale et Analytique, Université de Lausanne, BCH, CH-1015 Lausanne, Switzerland.
Abstract:The development of a stopped-flow instrument that operates over a temperature range of -40 to +100 °C and up to 200 MPa is described. The system has been designed so that measurements can be performed in absorbance and fluorescence modes simultaneously, without dismantling the unit. It can easily be combined with an optical system of a conventional ambient pressure setup by using light guides. Optimum optical performance and a wide operating wavelength range (220-850 nm) are achieved as the light is not passing through the pressurizing fluid. A special design for the pistons has been developed; thus, the apparatus has proven to be leak-free, even under extreme conditions (high pressure, low temperature, various solvents). The dead time of the system is found to be less than 2 ms at 298 K and is pressure independent up to 200 MPa. We examined the kinetics for the formation of the Mg(2+)-8-hydroxyquinoline chelate in aqueous solutions at pH 8.0 in order to develop a convenient alternative test method for high-pressure stopped-flow spectrometers with absorption and fluorescence detection.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号