首页 | 本学科首页   官方微博 | 高级检索  
     


Method of Monte Carlo simulation verification in hadron therapy with non-tissue equivalent detectors
Authors:Rosenfeld Anatoly  Wroe Andrew  Carolan Martin  Cornelius Iwan
Affiliation:Centre for Medical Radiation Physics, University of Wollongong, Wollongong NSW 2522, Australia. anatoly@uow.edu.au
Abstract:In hadron therapy the spectra of secondary particles can be very broad in type and energy. The most accurate calculations of tissue equivalent (TE) absorbed dose and biological effect can be achieved using Monte Carlo (MC) simulations followed by the application of an appropriate radiobiological model. The verification of MC simulations is therefore an important quality assurance (QA) issue in dose planning. We propose a method of verification for MC dose calculations based on measurements of either the integral absorbed dose or the spectra of deposited energies from single secondary particles in non-TE material detectors embedded in a target of interest (phantom). This method was tested in boron neutron capture therapy and fast neutron therapy beams.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号