首页 | 本学科首页   官方微博 | 高级检索  
     

快速馏程分析仪适用范围和准确性的探讨
引用本文:杨载松.快速馏程分析仪适用范围和准确性的探讨[J].广石化科技,2006(1):38-41.
作者姓名:杨载松
摘    要:介绍了快速馏程分析的原理,并做了大量试验,试验结果与GB/T6536-1997的进行对比,证明了该方法分析速度快(8~12min/次),适用测试的样品种类多,结果准确,其重复性相对偏差小于0.43%,可用于中控分析。

关 键 词:快速  馏程  适用性  准确性
收稿时间:2006-01-17

PROBE OF ACCURACY AND APPLICABLE SCOPE OF RAPID DISTILLATION ANALYZER
Authors:Yang Zaisong of Laboratory
Abstract:The principles of rapid distillation analyzing is introduced. A lot of analyses by rapid distillation analyzing were done, and then their results were compared with those whose analyzing were done according to GB/T 6536 - 1997. The comparison result shows that the time spent in each of the analyzing varies between 8 to 12 minutes, there are more samples which can be tested by rapid distillation analyzer, the analysis result is correct, and the relative standard deviation of repeatability is less than 0. 43%. Now it is clear that the rapid distillation analyzer is suitable for intermediate process control.
Keywords:rapid distillation range mobility accuracy
本文献已被 维普 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏