D-T中子源束斑尺寸测量技术及应用
引用本文:张凯,侯龙,陈红涛,鲍杰,苏明,阮念寿,赵芳. D-T中子源束斑尺寸测量技术及应用[J]. 原子能科学技术, 2020, 54(12): 2289-2295. DOI: 10.7538/yzk.2020.54.12.2289
作者姓名:张凯  侯龙  陈红涛  鲍杰  苏明  阮念寿  赵芳
作者单位:中国原子能科学研究院 核数据重点实验室,北京102413;中国工程物理研究院 材料研究所,四川 绵阳621000
摘    要:D-T中子发生器作为高能中子源,已用于特种核材料的有源探测技术中,利用时间符合法可实现被测物体的多模式成像,而中子源束斑尺寸是影响成像位置分辨的一个重要的因素。因此,结合D-T反应的特点和实际应用环境,开发了n-α关联符合测量确定中子源束斑尺寸的方法。利用该方法对小型移动中子发生器的束斑尺寸进行了测量,获得的束斑尺寸为(2.8±0.9) mm,与用CCD相机直接观测得到的约3 mm的测量结果一致,证明了该方法测量束斑尺寸的可行性。该方法也可用于辅助D-T中子源调束和关联粒子成像实验过程中的束斑监测。

关 键 词:D-T中子源  束斑尺寸测量  关联粒子成像  位置灵敏探测器

Beam Spot Size Measurement Technology and Application for D-T Neutron Source
ZHANG Kai,HOU Long,CHEN Hongtao,BAO Jie,SU Ming,RUAN Nianshou,ZHAO Fang. Beam Spot Size Measurement Technology and Application for D-T Neutron Source[J]. Atomic Energy Science and Technology, 2020, 54(12): 2289-2295. DOI: 10.7538/yzk.2020.54.12.2289
Authors:ZHANG Kai  HOU Long  CHEN Hongtao  BAO Jie  SU Ming  RUAN Nianshou  ZHAO Fang
Affiliation:Key Laboratory of Nuclear Data, China Institute of Atomic Energy, Beijing 102413, China;Institute of Special Materials, China Academy of Engineering Physics, Mianyang 621000, China
Abstract:As a high-energy neutron source, the D-T neutron generator has been used in the active detection technology of special nuclear materials. The time coincidence method can be used to achieve multi-mode imaging of the measured object, and the beam spot size of the neutron source is a very important factor that affects the imaging position resolution. Therefore, combined with the characteristics of D-T reaction and practical application environment, a method for n-α correlation coincidence measurement to determine the beam spot size of neutron source was developed. This method was used to measure the beam spot size of a small mobile neutron generator. The beam spot size obtained is (2.8±0.9) mm, which is consistent with the measurement result of about 3 mm obtained by direct observation with CCD camera, which proves that this method is feasible. This method can also be used to assist D-T neutron source beam tuning and beam spot size monitoring in associated particle imaging experiments.
Keywords:D-T neutron source   beam spot size measurement   associated particle imaging   position sensitive detector