首页 | 本学科首页   官方微博 | 高级检索  
     

一种变温环境下水听器灵敏度校准系统的设计
引用本文:陈思强,李雪健,王月兵,赵鹏.一种变温环境下水听器灵敏度校准系统的设计[J].计量学报,2023,44(2):238-245.
作者姓名:陈思强  李雪健  王月兵  赵鹏
作者单位:1. 中国计量大学 计量测试工程学院,浙江 杭州 310018
2. 国家管网集团西气东输公司 南京计量研究中心,江苏 南京 210046
基金项目:国家管网集团科技项目(CLZB202108)
摘    要:针对缺少在变温条件下校准水听器灵敏度方法的问题,设计了一套基于动圈换能器声管的变温水听器灵敏度校准系统。建立模型分析声管内的声压分布,并计算了变温后对灵敏度校准结果的补偿量;利用Comsol仿真不同声管厚度对声场的影响,分析了底面非均匀振动和不同尺寸的水听器对校准结果的影响,确定了合适的声管尺寸和测量方法。分别在常温与变温条件下对不同尺寸的水听器进行校准并将校准结果与标准值比较。结果表明,此系统在常温下的校准结果与标准值的偏差低于0.5 dB,变温条件下的测量结果与标准值偏差低于0.6 dB,整体系统的扩展不确定度约为1.0 dB。

关 键 词:计量学  水听器  动圈换能器  灵敏度  校准  变温环境
收稿时间:2021-12-27

Design of a Hydrophone Sensitivity Calibration System in Variable Temperature Environment
CHEN Si-qiang,LI Xue-jian,WANG Yue-bing,ZHAO Peng.Design of a Hydrophone Sensitivity Calibration System in Variable Temperature Environment[J].Acta Metrologica Sinica,2023,44(2):238-245.
Authors:CHEN Si-qiang  LI Xue-jian  WANG Yue-bing  ZHAO Peng
Affiliation:1. College of Metrology and Measurement Engineering, China Jiliang University, Hangzhou, Zhejiang 310018, China
2. Nanjing Metrology Research Center, PipeChina West East Gas Pipeline Company, Nanjing, Jiangsu 210046, China
Abstract:Aiming at the lack of a method for calibrating the sensitivity of hydrophones under variable temperature conditions, a temperature-variable hydrophone sensitivity calibration system based on the sound tube of the moving coil transducer was designed. A model was established to analyze the sound pressure distribution in the sound tube, and the compensation amount for the sensitivity calibration result after changing temperature was calculated; Comsol was used to simulate the influence of different sound tube thicknesses on the sound field, and the non-uniform vibration of the bottom surface and the effect of different sizes of hydrophones on the calibration were analyzed. Influence of the results, the appropriate acoustic tube size and measurement method were determined. The hydrophones of different sizes were calibrated under normal and variable temperature conditions and the calibration results were compared with the standard values. The results show that the deviation of the calibration result of this system from the standard value is less than 0.5dB at normal temperature, and the deviation of the measurement result from the standard value under the condition of variable temperature is less than 0.6dB, and the expanded uncertainty of the whole system is about 1.0dB.
Keywords:metrology  hydrophone  moving-coil transducer  sensitivity  calibrate  variable temperature environment  
点击此处可从《计量学报》浏览原始摘要信息
点击此处可从《计量学报》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号