高准确度水三相点容器
引用本文:孟哲健,包福兵,闫小克,尹招琴,凃程旭,滕俊恒.高准确度水三相点容器[J].计量学报,2023,44(2):195-202.
作者姓名:孟哲健  包福兵  闫小克  尹招琴  凃程旭  滕俊恒
作者单位:1.中国计量大学计量测试工程学院,浙江杭州310018
2.中国计量科学研究院,北京100029
基金项目:国家重点研发计划专项(2019YFC1408602;2017YFF0205901)
摘    要:水三相点的高精度复现及准确测量是保证国际温标ITS-90实施的关键。水三相点容器内高纯水的同位素组成会影响复现的水三相点温度值。为了提高水三相点复现水平,减小氢氧同位素的影响,研制了带有氢氧同位素分析的石英及硼硅玻璃高准确度水三相点容器。为了评价容器的性能,开展了硼硅玻璃和石英水三相点容器的比对。实验结果表明:同位素修正前,石英玻璃和硼硅玻璃水三相点容器复现的水三相点在0.058mK范围内一致;同位素修正之后,容器之间的差异在0.017mK范围内一致。采用高准确度水三相点容器复现水三相点的扩展不确定度为0.066mK(k=2)。

关 键 词:计量学  水三相点容器  同位素修正  ITS-90  不确定度  
收稿时间:2022-07-21

High Accuracy Triple Point of Water Cells
MENG Zhe-jian,BAO Fu-bing,YAN Xiao-ke,YIN Zhao-qin,TU Cheng-xu,TENG Jun-heng.High Accuracy Triple Point of Water Cells[J].Acta Metrologica Sinica,2023,44(2):195-202.
Authors:MENG Zhe-jian  BAO Fu-bing  YAN Xiao-ke  YIN Zhao-qin  TU Cheng-xu  TENG Jun-heng
Affiliation:1. College of Metrology and Measurement Engineering, China Jiliang University, Zhejiang,Hangzhou 310018, China
2. National Institute of Metrology, Beijing 100029, China
Abstract:The high-precision realization and accurate measurement of triple point of water (TPW) are the keys to ensure the implementation of the International Temperature Scale of 1990 (ITS-90). The isotopic compositions of the high purity water in the TPW cells affect the temperature of TPW. In order to improve the TPW realization and reduce the influence of hydrogen and oxygen isotopes, quartz and borosilicate glass high-accuracy TPW cells with hydrogen and oxygen isotopic analyses were developed. In order to evaluate the performance of the cells, comparisons between the borosilicate glass cells and the quartz cells were carried out. The experimental results show that, before the isotopic corrections, the temperature differences between the quartz and borosilicate glass high-accuracy TPW cells are within 0.058mK. After application of the corrections, the temperature differences are within 0.017mK. The expanded uncertainty of the realization of TPW using high-accuracy TPW cells is 0.066mK (k=2).
Keywords:metrology  triple point of water cell  isotopic correction  ITS-90  uncertainty  
点击此处可从《计量学报》浏览原始摘要信息
点击此处可从《计量学报》下载全文