首页 | 本学科首页   官方微博 | 高级检索  
     


Noninvasive thermometry assisted by a dual-function ultrasound transducer for mild hyperthermia
Authors:Lai Chun-Yen  Kruse Dustin E  Caskey Charles F  Stephens Douglas N  Sutcliffe Patrick L  Ferrara Katherine W
Affiliation:University of California at Davis, Department of Biomedical Engineering, Davis, CA, USA.
Abstract:Mild hyperthermia is increasingly important for the activation of temperature-sensitive drug delivery vehicles. Noninvasive ultrasound thermometry based on a 2-D speckle tracking algorithm was examined in this study. Here, a commercial ultrasound scanner, a customized co-linear array transducer, and a controlling PC system were used to generate mild hyperthermia. Because the co-linear array transducer is capable of both therapy and imaging at widely separated frequencies, RF image frames were acquired during therapeutic insonation and then exported for off-line analysis. For in vivo studies in a mouse model, before temperature estimation, motion correction was applied between a reference RF frame and subsequent RF frames. Both in vitro and in vivo experiments were examined; in the in vitro and in vivo studies, the average temperature error had a standard deviation of 0.7°C and 0.8°C, respectively. The application of motion correction improved the accuracy of temperature estimation, where the error range was 1.9 to 4.5°C without correction compared with 1.1 to 1.0°C following correction. This study demonstrates the feasibility of combining therapy and monitoring using a commercial system. In the future, real-time temperature estimation will be incorporated into this system.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏