An integrated system for real-time detection of stiff masses with a single compression |
| |
Authors: | Fahmy Ahmed S Krieger Axel Osman Nael F |
| |
Affiliation: | Electrical and Computer Engineering Department, The Johns Hopkins University, Baltimore, MD 21218, USA. fahmy@jhu.edu |
| |
Abstract: | An integrated system is proposed for real-time imaging of tissue deformation in response to externally applied mechanical compression. The purpose of the system is to noninvasively detect tissue lesions based on their stiffness. The method includes applying external compression to the examined subject and acquiring magnetic resonance strain-encoded images using a real-time pulse sequence. In the proposed system, only a single compression is necessary to obtain a two-dimensional map of tissue deformation in a slice, eliminating the need for a complicated compressing device. The system was built and validated using phantom experiments. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|