首页 | 本学科首页   官方微博 | 高级检索  
     


Gradient-based 2-D/3-D rigid registration of fluoroscopic X-ray to CT
Authors:Livyatan Harel  Yaniv Ziv  Joskowicz Leo
Affiliation:School of Engineering and Computer Science, The Hebrew University of Jerusalem, Jerusalem 91904, Israel.
Abstract:We present a gradient-based method for rigid registration of a patient preoperative computed tomography (CT) to its intraoperative situation with a few fluoroscopic X-ray images obtained with a tracked C-arm. The method is noninvasive, anatomy-based, requires simple user interaction, and includes validation. It is generic and easily customizable for a variety of routine clinical uses in orthopaedic surgery. Gradient-based registration consists of three steps: 1) initial pose estimation; 2) coarse geometry-based registration on bone contours, and; 3) fine gradient projection registration (GPR) on edge pixels. It optimizes speed, accuracy, and robustness. Its novelty resides in using volume gradients to eliminate outliers and foreign objects in the fluoroscopic X-ray images, in speeding up computation, and in achieving higher accuracy. It overcomes the drawbacks of intensity-based methods, which are slow and have a limited convergence range, and of geometry-based methods, which depend on the image segmentation quality. Our simulated, in vitro, and cadaver experiments on a human pelvis CT, dry vertebra, dry femur, fresh lamb hip, and human pelvis under realistic conditions show a mean 0.5-1.7 mm (0.5-2.6 mm maximum) target registration accuracy.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号