首页 | 本学科首页   官方微博 | 高级检索  
     


X-ray imaging with “edge-on” microchannel plate detector: first experimental results
Authors:Polad M Shikhaliev  Sabee Molloi
Affiliation:

Department of Radiological Sciences, Imaging Physics Laboratory, University of California, Medical Sciences I, B-140, Irvine, CA 92697, USA

Abstract:A novel scanning slit X-ray imaging system based on an “edge-on” microchannel plate detector was developed and tested. Images were acquired at 50 kV(p) X-ray tube voltage with a limiting spatial resolution of 7 lp/mm. The pixel noise was measured to be 0.3 count/pixel/s for a 50×70 μm2 pixel size. This photon counting detector can be considered to be virtually noise free.
Keywords:X-ray imaging  MCP detector  “Edge-on” method  Scanning slit
本文献已被 ScienceDirect 等数据库收录!