Quantitating Fluorescence Intensity From Fluorophores: Practical Use of MESF Values
Authors:Lili Wang  Adolfas K Gaigalas  Fatima Abbasi  Gerald E Marti  Robert F Vogt  Abe Schwartz
Affiliation:National Institute of Standards and Technology, Gaithersburg, MD 20899-8312;Center for Biologics Evaluation and Research, U.S. Food and Drug Administration, NIH Building 29B, Room 2NN08, Bethesda, MD 20892;Division of Laboratory Sciences, CDC, Mailstop F19, 4770 Buford Highway, Atlanta, GA 30341;Center for Quantitative Cytometry, PO Box 194344, San Juan, Puerto Rico, 00919
Abstract:The present work uses fluorescein as the model fluorophore and points out critical steps in the use of MESF (Molecules of Equivalent Soluble Fluorophores) values for quantitative flow cytometric measurements. It has been found that emission spectrum matching between a reference solution and an analyte and normalization by the corresponding extinction coefficient are required for quantifying fluorescence signals using flow cytometers. Because of the use of fluorescein, the pH value of the medium is also critical for accurate MESF assignments. Given that the emission spectrum shapes of microbead suspensions and stained biological cells are not significantly different, the percentage of error due to spectrum mismatch is estimated. We have also found that the emission spectrum of a microbead with a seven-methylene linker between the fluorescein and the bead surface (bead7) provides the best match with the spectra from biological cells. Therefore, bead7 is potentially a better calibration standard for flow cytometers than the existing one that is commercially available and used in the present study.
Keywords:emission spectrum matching  extinction coefficient  fluorescein  lymphocyte  MESF value  microbead  pH  quantitative flow cytometry
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号