毛细管波导流通池连续流动分析仪测定水中低浓度挥发酚
引用本文:曾次元,任嫣婵,赵鉴,叶秋明. 毛细管波导流通池连续流动分析仪测定水中低浓度挥发酚[J]. 分析仪器, 2011, 0(4): 49-52
作者姓名:曾次元  任嫣婵  赵鉴  叶秋明
作者单位:上海市供水调度监测中心,上海,200002
摘    要:采用配有1m毛细管波导流通池的连续流动分析仪(CFA)测定了水中的低浓度挥发酚,探讨了方法的检出限、准确度、精密度和水样加标回收率。结果显示,仪器方法的检出限为0.11μg/L,考核样测定平均值为4.41μg/L,与标准值4.55±0.36μg/L吻合,相对标准偏差为0.79%。测定实际水样,加标回收率在96%~104%之间。本法检出限低,准确性好,精密度高,适合水中低浓度挥发酚的检测。

关 键 词:毛细管波导流通池  连续流动分析仪(CFA)  挥发酚

Determination of trace volatile phenol in water by continuous flow analyzer with liquid waveguide capillary cell
Zeng Ciyuan,Ren Yanchan,ZhaoJian,Ye Qiuming. Determination of trace volatile phenol in water by continuous flow analyzer with liquid waveguide capillary cell[J]. Analytical Instrumentation, 2011, 0(4): 49-52
Authors:Zeng Ciyuan  Ren Yanchan  ZhaoJian  Ye Qiuming
Affiliation:Zeng Ciyuan,Ren Yanchan,Zhao Jian,Ye Qiuming(Shanghai Municipal Water Supply Dispatching and Monitoring Center,Shanghai,200002)
Abstract:Volatile phenol in water was determined by continuous flow analyzer(CFA) with a lm liquid waveguide capillary cell.The detection limit,precision,accuracy and recovery of standard addition were inverstigated.The results show that the detection limit is 0.11μg/L.The average value of quality control sample determined by this method is 4.41μg/L,which is in agreement with the standard value(4.55±0.36μg/L).The relative standard deviation is 0.79% and the recovery is 96%~104%.The method has low detection limit,good accuracy and high precision,so it is especially suitable for the determination of low concentration volatile phenol in water.