Investigation into calibration performance of small volume prover for hydrocarbon flow
Affiliation:1. Faculdade de Engenharia de Ilha Solteira - UNESP, Ilha Solteira, São Paulo, Brazil;2. Departamento de Engenharia Mecânica, Av. Brasil Sul, 55, 15385-000, Ilha Solteira, São Paulo, Brazil;3. Departamento de Engenharia Civil, Alameda Bahia, 550, 15385-000, Ilha Solteira, São Paulo, Brazil;4. Departamento de Engenharia Elétrica, Campus III, 15385-000, Ilha Solteira, São Paulo, Brazil
Abstract:Several kinds of commercial flowmeters, namely, Coriolis flowmeters, turbine meters, ultrasonic flowmeters, and positive displacement flowmeters, have been calibrated using the primary standard for hydrocarbon flow measurement in Japan (which is based on static and gravimetric methods with a flying start and finish) and a small volume prover (SVP) at the same calibration condition in order to investigate the performance of the SVP. The differences in calibration results for the mechanical flowmeters between the primary standard and the SVP apparently depend on the flow rate, although the results show agreement within 0.04%. The computer-based flowmeters, which have a time delay in the output pulse signal, indicated larger differences due to the effect of the sudden flow rate change caused by the proving action of the SVP at larger flow damping times.
Keywords:Flowmeter  Calibration  Small volume prover  Uncertainty