首页 | 本学科首页   官方微博 | 高级检索  
     


Influence of electrosorption, solvent, temperature, and ion polarity on the performance of LC-ESI-MS using graphitic carbon for acidic oligosaccharides
Authors:Pabst Martin  Altmann Friedrich
Affiliation:Department of Chemistry, University of Natural Resources and Applied Life Sciences (BOKU), 1190 Vienna, Austria.
Abstract:Porous graphitic carbon (PGC) emerges as an ideal stationary phase for LC-ESI-MS of complex oligosaccharides. Therefore, we studied the factors influencing detection and elution of charged oligosaccharides from PGC columns coupled to an ESI source. Electrosorption by the carbon surface leads to total retention of very acidic glycans on instruments where voltage is applied to the spray needle. This problem can be eliminated by thorough electrical grounding. A point of general importance is the influence of ionic strength on the elution and peak shape of glycans containing several carboxylic acid groups in the form of sialic acids or uronic acids. Solvent pH had a marginal effect on the ionization efficiency in both ion polarities, but the content of organic solvent strongly influenced signal intensity of acidic glycans in the negative mode. As a consequence, detection in the positive ion mode appears preferable when neutral and charged glycans shall be quantitated in the same sample. While retention of neutral glycans is not affected by pH, sialylated species are retained somewhat stronger at acidic pH resulting in a larger spread of the entire elution range of N-glycans. Remarkably, retention of glycans on PGC increased at higher temperatures.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号