Imaging of immunolabeled membrane receptors in uncoated sem specimens
Authors:U Heinzmann  A Reininger  R Autrata  H Hfler
Affiliation:U. Heinzmann,A. Reininger,R. Autrata,H. Höfler
Abstract:Epidermal growth factor receptors (EGFR) were labeled with 10 nm immunogold and examined on uncoated specimens of A431 human epidermoid carcinoma cells. A field emission gun and a high-sensitivity YAG ring detector were used to demonstrate the affinity labeling simultaneously in the secondary-electron (SE) and backscattered-electron (BSE) modes with a low accelerating voltage (Vo). At Vo=2kV, the SE and BSE signals were too weak to identify all markers, while at Vo=3–7 kV labeling was observed unambiguously in both the SE and BSE modes with smaller and higher working distances. Increasing the Vo to above 7 kV sometimes provokes instability of the specimens. A Vo of ? 10 kV produces charging artifacts in the SE image, but permits a BSE image of the gold markers providing additional topographic information. In conclusion, immunogold labeling can be used with good results for uncoated specimens.
Keywords:immunogold labeling  epidermal growth factor receptors  low-voltage scanning electron microscopy  secondary electron  backscattered electrons  uncoated biological specimens