Gas chromatograph/isotope dilution mass spectrometric analysis of airborne benzo |
| |
Authors: | R Luniewski B Kanabe L Nasri P Hubber L Au P Yang N Buonocore |
| |
Affiliation: | Laboratory Services Branch, Ontario Ministry of Environment and Energy, Canada. |
| |
Abstract: | A critical evaluation has been made of the gas chromatography/isotope dilution mass spectrometry (GC/IDMS) analysis of airborne benzoa]pyrene (Ba]p) using 2H-(deuterium, D) and 13C-labeled benzoa]pyrene and a low resolution mass spectrometer. Laboratory quality control and field samples were used to determine the suitability of these two isotopically labeled Ba]p as surrogates for GC/IDMS Ba]p analysis. With method spike QC samples, recovery of the Ba]p was compared with the recoveries of D- and 13C-labeled Ba]p to validate the assumption that these two isotopically labeled Ba]p analogs could be used as method surrogates for IDMS Ba]p analyses. Two collocated Anderson Hi-Vol samplers were used to carry out a field study in which total Ba]p loadings from one sampler were correlated to the second sampler, before and after correcting the total Ba]p loadings against the recovery of the D- and 13C-labeled Ba]p. It is shown that a low-resolution mass spectrometer can be used in GC/IDMS Ba]p analysis to achieve results of high precision and accuracy. Depending on data quality objectives, D-labeled Ba]p could be used for IDMS Ba]p analysis, but is not recommended. |
| |
Keywords: | |