Measurement of diagnostic electronic linear arrays by miniature hydrophone scanning |
| |
Authors: | Ide M Ohdaira E |
| |
Affiliation: | Dept. of Electron. and Commun. Eng., Musashi Inst. of Technol., Tokyo. |
| |
Abstract: | Ultrasonic pressures, intensities and power were computed using a calibrated miniature hydrophone scanning system. To accomplish these measurements, the electronically scanned beam was stopped, and raster scans were made with the hydrophone over a number of planes perpendicular to the beam axis. Ultrasonic power also was measured using a radiation force balance, and the results compared well with those obtained using the hydrophone. Spatial average-temporal average intensities for the stopped beam ranged from 0.13 to 3.71 mW/cm(2 ) (for a -40-dB beam area), and spatial peak-temporal peak intensities varied from 8.31 to 45.2 W/cm(2). Such measurements are important in setting equipment safety standards, in encouraging minimum use of ultrasonic exposures, and in quantifying exposure levels that may result in harmful biological effects. |
| |